Changsha Panran Technology Co., Ltd.

PANRAN is committed to providing you with comprehensive professional temperature and pressure metrology solutions!

Manufacturer from China
Active Member
5 Years
Home / Products / Liquid Calibration Baths /

Infrared ISO9001 150degree Liquid Calibration Baths

Contact Now
Changsha Panran Technology Co., Ltd.
City:changsha
Province/State:hunan
Country/Region:china
Contact Person:PANRAN
Contact Now

Infrared ISO9001 150degree Liquid Calibration Baths

Ask Latest Price
Video Channel
Brand Name :PANRAN
Model Number :PR565
Certification :ISO9001
Place of Origin :Tai'an,China
MOQ :1 set
Payment Terms :TT, LC, Visa, Western Union, Alipay, Trade assurance, E-checking
Supply Ability :10 sets/month
Delivery Time :20~35 working days
Packaging Details :Standard export Packing (aluminium alloy, laminated wood, carton)
Price :negotiable
Emissivity(8~14 μm wave length) :≥0.999 and ≥0.997
Diameter of the hole :10mm and 60mm
Max immersion depth :150mm and 300mm
Flange diameter :130mm
Temperature range :-10~150℃
Accuracy :0.1℃+0.1%RD
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

The Infrared Ear Thermometer And Forehead Thermometer Calibration System

PanRan Measurement & Control provides a comprehensive infrared ear thermometer and infrared forehead thermometer calibration solution.

The infrared ear thermometer and forehead thermometer calibration system consists of three parts:

Part1. Black-body radiation cavity, high-emissivity black-body radiation cavity is a key component required for the calibration of infrared ear thermometers and forehead thermometers. Its structure and the quality of the internal coating have a significant impact on the calibration results.

Part2. Temperature source--Liquid constant temperature device, used to place and immerse the black body radiation cavity, so that each surface of the radiation cavity has excellent temperature uniformity and temperature fluctuation.

Part3. Temperature standard, used to measure the temperature of the medium in the liquid thermostat.

Infrared ISO9001 150degree Liquid Calibration Baths

Part1. Black-body radiation cavity

There are two types of black body radiation chambers, which are used to calibrate infrared ear thermometers and infrared forehead thermometers. The black body cavity is gold-plated on the outside and has a high-emissivity coating inside. Requirements to meet the calibration requirements of most infrared ear thermometers and infrared forehead thermometers.

Item

HC1656012

For infrared ear thermometer calibration

HC1686045

For infrared forehead thermometer calibration

Emissivity(8~14 μm wave length) ≥0.999 ≥0.997
Diameter of the hole 10mm 60mm
Max immersion depth 150mm 300mm
Flange diameter 130mm

Infrared ISO9001 150degree Liquid Calibration BathsInfrared ISO9001 150degree Liquid Calibration BathsInfrared ISO9001 150degree Liquid Calibration Baths

Part2. Temperature source--the liquid constant temperature device

The liquid constant temperature device can choose two types of products, the PR560B infrared thermometer calibration thermostat or the PR532-N10 refrigeration thermostat, both of which have excellent temperature stability and temperature uniformity. Among them, the volume of the thermostat used for the calibration of the PR560B infrared thermometer is only 1/2 of that of an ordinary thermostat, which is convenient to move, transport, or transform it into a vehicle-mounted calibration device.

Items

PR560B

infrared thermometer calibration thermostatic bath

PR532-N10

Cooling bath

Remarks
Temperature range 10~90℃ -10~150℃ Environment temp. 5℃~35℃
Accuracy

36℃,≤0.07℃

Full range,≤0.1℃

0.1℃+0.1%RD
Work medium Distilled water antifreeze
Resolution 0.001℃
Temperature uniformity 0.01℃

Full range

From 40mm from the bottom

Temperature stability ≤0.005℃/1min ≤0.01℃/10min 20 minutes after reaching the set temperature
Power supply 220VAC,50Hz,2KVA
Dimension 800mm×426mm×500mm(H×H×W)
Weight 60KG

Infrared ISO9001 150degree Liquid Calibration Baths

Note: If the customer already has a constant temperature device that can meet the calibration requirements, it can also be used directly.

Part3. Temperature standard

Option 1: In response to the calibration requirements of infrared thermometers, PanRan introduced the PR712A standard digital thermometer, with an annual change of better than 0.01 ° C over the full range. Compared with the PR710 and PR711 precision digital thermometers of the same series, it has a better built-in reference resistance, better temperature coefficient and long-term stability. At an ambient temperature of 10 to 35 ° C, its typical temperature coefficient is only 0.5 ppm / ° C.

Option 2: Traditional electrical measurement equipment + standard platinum resistance. The electrical measurement equipment in this solution can be configured with PR293 series nanovolt micro-ohm thermometer or PR291 series micro-ohm thermometer. Both series of products can meet the requirements of electrical thermometers related to infrared thermometers.

Items

PR712A

Standard digital thermometer

PR293 series

Nanovolt microohm thermometer

PR291 series

Microohm thermometer

Remarks
Description High-precision integrated thermometer,temp. sensor is wound typePT100,sensorφ5*400mm. Full-featured thermocouple and platinum resistance thermometer High-precision platinum resistance thermometer
Channel No. 1 2 or 5 2
Accuracy 0.01℃

Electricity:20ppm(RD)+2.5ppm(FS)

Temperature:36℃,≤0.008℃

The PR291 and PR293 thermometers use standard platinum resistance measurement functions.
Resolution 0.001℃ 0.0001℃
Temperature range -5℃~50℃ -200℃~660℃
Communication 2.4G wireless RS485
Battery power duration >1400h >6h PR712A power is AAA battery
Dimention(Body) 104×64×30mm 230×220×112mm
Weight 110g 2800g Including battery weight

Application:

The high precision cooling thermostatic bath is suitable for measuring, biochemical, petroleum, meteorology, energy, environmental protection, medicine and other departments and manufacturers of thermometers, temperature controllers, temperature sensors and other manufacturers to test and calibrate physical parameters. It can also provide constant temperature source for other experimental research work. Example 1. Second-class standard mercury thermometer, Forehead thermometers, Infrared Surface Thermometers, Ear Thermometers, beckman thermometer, industrial platinum thermal resistance, standard copper-constantan thermocouple verification, etc.